# Displacement response = Nanometrics Trillium Compact + ED
# Seismometer sensitivity from manual 749.1 Vs/m
# ED sensitivity 1000000 counts/V
# System sensitivity 749.1x1000000 = 0.74910 counts/nm/s
# Normalization factor 8.184e+11 (from Trillium manual)
# std conts. => 0.74910 x 8.184e+11 = 6.1306344e+11
theoretical 1 instrument paz Kortstrom
6.1306344e+11
7
-0.03691 0.03712 0.0000000000e+00 0.0000000000e+00
-0.03691 -0.03712 0.0000000000e+00 0.0000000000e+00
-371.2 0.00000 0.0000000000e+00 0.0000000000e+00
-373.9 475.5 0.0000000000e+00 0.0000000000e+00
-373.9 -475.5 0.0000000000e+00 0.0000000000e+00
-588.4 1508 0.0000000000e+00 0.0000000000e+00
-588.4 -1508 0.0000000000e+00 0.0000000000e+00
4
-434.1 0.00000 0.0000000000e+00 0.0000000000e+00
0.000 0.00000 0.0000000000e+00 0.0000000000e+00
0.000 0.00000 0.0000000000e+00 0.0000000000e+00
0.000 0.00000 0.0000000000e+00 0.0000000000e+00