# Displacement response  Nanometrics compact 
# + Reftek 130
# Seismometer sensitivity 749.1  V/m/s   
# reftek sensitivity 1.589 muV/count = 629326.6 count/V 
# System sensitivity 749.1x629326.6 = 0.471 count/nm/s
# ncal 2*pi*1*0.471 = 2.95938 counts/nm
#  == 0.337909 nm/count @ ncalper=1s.
# Normalization factor 8.184e+11 (from Trillium manual)
# std conts. => 0.471 x 8.184e+11 x 2pi*1Hz = 2.421956821e+12
theoretical   1 instrument   paz    Kortstrom
2.421956821e+12
       7
-0.03691  0.03712   0.0000000000e+00 0.0000000000e+00
-0.03691 -0.03712   0.0000000000e+00 0.0000000000e+00
-371.2    0.00000   0.0000000000e+00 0.0000000000e+00
-373.9    475.5     0.0000000000e+00 0.0000000000e+00
-373.9   -475.5     0.0000000000e+00 0.0000000000e+00
-588.4    1508      0.0000000000e+00 0.0000000000e+00
-588.4   -1508      0.0000000000e+00 0.0000000000e+00
      4
-434.1   0.00000   0.0000000000e+00 0.0000000000e+00
0.000    0.00000   0.0000000000e+00 0.0000000000e+00
0.000    0.00000   0.0000000000e+00 0.0000000000e+00
0.000    0.00000   0.0000000000e+00 0.0000000000e+00