Newer
Older
# Displacement response = Nanometrics 120QA + Centaur
# Seismometer sensitivity from manual 1202 Vs/m
# ED sensitivity 800000 counts/V
# System sensitivity 1202x800000 = 0.9616 counts/nm/s
# ncal 2*pi*1*0.9616 = 6.041911 counts/nm
# == 0.165511 nm/count @ ncalper=1s.
# Normalization factor 8.318710E+17 (from Trillium manual)
# std conts. => 0.9616 x 8.318710E+17 = 7.999272e+17
theoretical 1 instrument paz Kortstrom
7.999272e+17
11
-3.661400e-02 3.705900e-02 0.000000E+00 0.000000E+00
-3.661400e-02 -3.705900e-02 0.000000E+00 0.000000E+00
-3.255000e+01 0.000000e+00 0.000000E+00 0.000000E+00
-1.420000e+02 0.000000e+00 0.000000E+00 0.000000E+00
-3.640000e+02 4.040000e+02 0.000000E+00 0.000000E+00
-3.640000e+02 -4.040000e+02 0.000000E+00 0.000000E+00
-1.260000e+03 0.000000e+00 0.000000E+00 0.000000E+00
-4.900000e+03 5.204000e+03 0.000000E+00 0.000000E+00
-4.900000e+03 -5.204000e+03 0.000000E+00 0.000000E+00
-7.100000e+03 1.700000e+03 0.000000E+00 0.000000E+00
-7.100000e+03 -1.700000e+03 0.000000E+00 0.000000E+00
7
0.000000e+00 0.000000e+00 0.000000E+00 0.000000E+00
0.000000e+00 0.000000e+00 0.000000E+00 0.000000E+00
0.000000e+00 0.000000e+00 0.000000E+00 0.000000E+00
-3.163000e+01 0.000000e+00 0.000000E+00 0.000000E+00
-1.600000e+02 0.000000e+00 0.000000E+00 0.000000E+00
-3.500000e+02 0.000000e+00 0.000000E+00 0.000000E+00
-3.177000e+03 0.000000e+00 0.000000E+00 0.000000E+00